COMPANY PRODUCT

310 Surface Roughness Gauge

Product description:

The performance advantage The roughness of metal and non-metallic surface Pocket-size economically price Large measuring range suitable for most materials Features external calibration at keyboard Opt
  • TIME®3221 Surface Roughness Tester
  • TIME®3200/3202  Handheld Surface Roughness Tester
  • HR 3100 Surface Roughness Tester
  • 310 Surface Roughness Gauge

Share With:

We"re here to help:
Easy ways to get the answers you need.
Send E-mailChat Now
The performance advantage
Ø The roughness of metal and non-metallic surface
Ø Pocket-size & economically price
Ø Large measuring range suitable for most materials
Ø Features external calibration at keyboard
Ø Optimized electric circuit design with transducer structure-design,high integrate power, driver and display. 
Ø Select freely with Ra,Rz, Rq, Rt parameters.
Ø It could test ex-circle, flat surface, conical surface and also test groove with length and width larger than 80*30mm.
 
 
Main Function:
v Shape used to pull aluminum mold design, durable, anti-electromagnetic interference ability significantly, in line with the design of the new trend.
v Using high-speed DSP processor for data processing and calculation, measurement and computation speed is greatly improved.
v Liquid crystal display wide temperature using OLED color display, popular high brightness, no visual angle,, is suitable for various occasions.
v The lithium ion rechargeable battery, can work for a long time, no memory effect, can while charging, charging time is short, long battery life.
v Charging and communication using the common USB interface. Using a dedicated charger or computer USB port for charging, convenient and quick.
v Dot matrix liquid crystal display, interface prompt information rich.
v Real-time monitoring of lithium battery power and display, and to remind users to charge with charging progress indication.
v Automatic shutdown function and low power design of software and hardware of the instrument with long working hours, suitable for field use.
 
Specification:
u Testing parameters(µm ):Ra、Rz、Rq、Rt
u Stroke length(mm):6
u Sample length(mm):0.25 ,0.80 ,2.50
u Evaluation length(mm):1.25 ,4.0 
u Measuring range(µm ): Ra:  0.05 ~ 15.0  Rz:  0.1 ~ 50
u Error:±15%
u Variability: <12%
u Transducer contacting pin round diameter and angle:
u Contacting pin round diameter:10 µm±1 µm
u Battery:3.7V Li
u Size:106 mm×70 mm×24 mm
u Weight:200g
u Sensor needle static force and its rate of change:
 
    Stylus static force: ≤ 0.016N
 
    Force measuring rate of change: ≤ 800N/m
u Working conditions:
Temperature:  -20℃~40℃
Relative humidity:   <  90%
No shake and coercivity around
 
Schedule: sampling length selection recommendation form:
 
Ra (µm) Rz (µm) The sampling length λc(mm)
>5~10 >20~40 2.5
>2.5~5 >10~20
>1.25~2.5 >6.3~10 0.8
>0.63~1.25 >3.2~6.3
>0.32~0.63 >1.6~3.2
>0.25~0.32 >1.25~1.6 0.25
>0.20~0.25 >1.0~1.25
>0.16~0.20 >0.8~1.0
>0.125~0.16 >0.63~0.8
>0.1~0.125 >0.5~0.63
>0.08~0.1 >0.4~0.5
>0.063~0.08 >0.32~0.4
>0.05~0.063 >0.25~0.32
>0.04~0.05 >0.2~0.25
>0.032~0.04 >0.16~0.2
>0.025~0.032 >0.125~0.16
>0.02~0.025 >0.1~0.125
 
 
Standard configuration:
 
Standard configuration
 
  Name QTY
1 KR100 host  1
2 reticle calibration block 1
3 bracket 1
4 Charger 1
5 data line 1
6 manual 1
7 warranty card 1
8 certificate 1



admin@hssdtest.com

+86-15910081986